EDAN Instruments, Inc | Patient Monitor M50


M50

Patient Monitor - EDAN Instrumetnts, Inc. - M50
The new generation of EDAN patient monitors M50 provides perfect transportation solution. The multi-functional all-in-one parameter module and new application platform greatly enhance the efficiency of full clinical parameter monitoring for patient transfer.

Configuration
Model Standard Optional
M50 3/5-lead ECG, EDAN SpO2, RESP, NIBP, 2-Temp, PR Touchscreen, Thermal Recorder, WLAN accessory, Nellcor-SpO2, 2-IBP, Respironics CO2 (for intubation and non-intubation application), 12V DC input

Features
8.4″ color TFT display with maximum 11 waveforms
Pacemaker detection, electrosurgical interference proof 
Defibrillation protection and defibrillation synchronization 
Full touchscreen enables intuitive operation by clicking on the specific parameters or the waveform in real time, standard configuration includes the navigation knob 
OxyCRG available to judge the respiration and circulation function for neonates 
Large font display
Arrhythmia and S-T segment analysis
Built-in rechargeable Li-ion battery 
Nurse Call function and bi-directional communication with MFM-CMS central station 
SD card slot enables memory extension for large data storage 
USB and serial ports enables abundant future upgrades 
VGA output and analog output 
Complete and flexible mounting solution for different scenarios

Specification
Safety Standards CE Marking in accordance to Council
Directive 2007/47/EC concerning Medical Devices
IEC 60601-1+A1+A2;IEC 60601-1-2+A1; 
Physical Specification
Device dimension 260 mm (L) x 140 mm (W) x 205 mm (H)
Weight Standard Configuration 3.6 kg 
Display
Color TFT LCD 8.4 Touchscreen Optional
Resolution 800 X 600 dpi
Traces Displayed up to 8
Waveforms Displayed up to 11
Various Working Interface Selectable Standard Monitoring Display
Large Font Intensive Care Display
Trend Graph/ Monitoring Co-Display
Bed to Bed view Display (Optional)
OxyCRG Dynamic View Display
Drug Dose Calculation Interface
Sweep Speed 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s 
Operation Environment
Ambient Temperature -20 - 55 (-4 -131 F)
Humidity 15%-95% non-condensing 
Power Supply
External Power Supply 100-240V AC, 50/60HZ Pmax=80VA
Internal Battery Power Supply
Type Rechargeable Li-ion
Voltage 14.8 V DC
Capacitance 2100 mAh (4200 mAh optional)
Battery working period up to 6 hours
Recharging time < 150 minutes (2100mAh)
< 360 minutes (4200mAh)) 
Resp Method Trans-thoracic impedance
Operation mode Auto/ Manual
RR Measurement range
Adult 0~120 rPM
Neonate/Pediatric 0~150 rPM
Resolution 1rPM
Apnea alarm threshold 10s,15s,20s (default),25s,30s,35s,40s
Alarm Yes, audible and visual alarm, alarm events recallable
Band width 0.2-2.5Hz (-3dB)
Sweep speed 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
ECG Lead type 5-lead and 3-lead selectable
3 leadwire cable RA; LA; LL or R; L; F
5 leadwire cable RA; LA; RL; LL; V or R; L; N; F; C
Lead selection 3-lead
Gain selection x0.125;x0.25; x0.5; x1; x2; x4; auto
Sweep speed 6.25mm/s,12.5mm/s, 25mm/s, 50mm/s
ECG HR Range
Adult/Pediatric 15-300bpm;
Neonate 15-350bpm
Resolution & accuracy 1bpm or 1%, whichever is greater
Filter
Diagnostic mode 0.05-150Hz
Monitoring mode 0.5-40 Hz
Surgical mode 1-20Hz
Protection Withstand 5000VAC/50Hz voltage in isolation against Defibrillation and electrosurgical interference
ST-Segment Detection
Measurement range -2.0 mV~2.0mV
Alarm range -2.0 mV~2.0mV
ST-Segment Arrhythmia analysis and catergorization  Yes
ARR Alarm Audible and visual alarm, alarm events recallable
Pace maker detection Yes, and 5 types abnormal status detectable
IEC 60601-2-25; AAMI EC 11/EC 13
IEC 60601-2-27 
NIBP Method Automatic Oscillometric
Operation modes Manual/Automatic/Continuous
Auto measurement time interval Adjustable 1/2/3/4/5/10/15/30/60/90/120/240/480 Minutes
Measurement unit mmHg/kPa selectable
Measurement types Systolic, Diastolic, Mean
Pressure range for Adults Systolic
Diastolic 10 - 215 mmHg
Mean 20 - 235 mmHg
Pressure range for Pediatrics Systolic
Diastolic 10 - 150 mmHg
Mean 20-165 mmHg
Pressure range for Neonates Systolic
Diastolic 10 - 100 mmHg
Mean 20 - 110 mmHg
Leak test and pressure auto calibration Yes
Over-pressure protection Dual Safety protection
Resolution 1mmHg
Accuracy Max mean error 5 mmHg
Max standard deviation  8 mmHg
Alarm Systolic, Diastolic, Mean
PR from NIBP Measurement 40~240 bpm
Resolution 1 bpm
Accuracy 3bpm or 3% whichever is greater
Leak test and pressure auto calibration Yes
IEC 60601-2-30 
EDAN SpO2 Measurement & Alarm Range 0 - 100%
Resolution 1%; Accuracy:  2% (70-100%, Adult/Pediatric);
3% (70-100%, Neonate)
PR Measurement and Alarm Range 30 - 300bpm
Resoluton 1 bpm;
Accuracy 3bpm
Refresh rate 2s
ISO 9919 
Nellcor SpO2 (optional) Measurement & Alarm Range 0 - 100%
Resolution 1%;
Accuracy 2-3% (70-100%, Adult/Pediatric);
3-3.5% (70-100%, Neonate)
PR Measurement and Alarm Range 20 - 300 bpm
Resoluton 1 bpm;
Accuracy 3bpm (depends on probe)
Refresh Rate 1s
ISO 9919 
Temperature (2 Channels, 1 probe by default) Measurement /Alarm range 0~50 (32-122 F)
Probe type YSI (B Series) and CF-FI
Resolution 0.1
Accuracy 0.1 (without probe)
Channel Dual-channel. Provide T1; T2; T
IEC 12470-4 
Quick Temperature (Optional) Measurement /Alarm range 25~45
Probe type Oral/ Axillary / Rectal
Resolution 0.1
Accuracy 0.1 (without probe)
Typical measurement time <15s
IEC 12470-4 
IBP (2-Channels, optional) Measuremed Pressure: ART, PA, CVP, RAP, LAP, ICP, P1,P2
Measurement range: -50 - 300 mmHg; Resolution 1 mmHg
Accuracy 2% or 1 mmHg, whichever is greater (without probe)
Sensitivity 5 V/V/mmHg;Impedance range:300-3000
IEC 60601-2-34 
CO2 (Mainstream and Sidestream, optional) By Philips Respironics CAPNOSTAT 5 & LoFlo Technology
Range 0~ 150mmHg
Accuracy 2% 0 ~ 40mmHg,
5% 41~70mmHg
8% 71~100mmHg
10% 101~150 mmHg
AwRR Accuracy 1 rpm
Convenient design for intubbated and non-intubated applciations
Possible to work at low sample flow rate 50ml / minute
Detailed specification refer to the user manual of Respronics
ISO 21647 
Recorder (optional) Built-in, direct thermal pixel array recorder
Up to 3 channels printing 1,2, 3 channels selectable
Print speed 25mm/s, 50mm/s
Paper width 50 mm/Record width:48mm 
I/O Interface) 2-USB Ports
SD Card Socket
RJ-45 Ethernet Port. IEEE 802.3
VGA Output
Analog & Nurse Call output
Defibrillation Synchronization output
WLAN Access Point 802.11g 54Mbps (optional 