3-part differentiation of WBC; 19 parameters+3 histograms
2 Counting modes: whole blood and prediluted
Throughput: 60 samples per hour
Automatic diluting, lyzing, mixing, rinsing and clog-clearing
Up to 20000 sample results (including histograms) can be stored
Built-in thermal recorder, optional external printer optional external printer and barcode scanner
Large color LCD display
Technical Specifications
Parameters
WBC, Lymph#, Mid#, Gran#, Lymph%, Mid%, Gran%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PDW, PCT and Histogram for WBC, RBC, PLT
Principles
Electrical resistance for counting and cyanide free method for hemoglobin
Performance
| Parameter |
Linearity Range |
Precision(CV%) |
| WBC(109/L) |
0.0-99.9 |
2.5 (7.0-15.0) |
| RBC(10¹²/L) |
0.00-9.99 |
2 (3.50-6.00) |
| HGB (g/L) |
0-300 |
1.5 (110-180) |
| MCV (fL) |
40.0-150.0 |
0.5 (80.0-110.0) |
| PLT(109/L) |
10-999 |
5 (200-500) |
Sample Volume
Prediluted 20 μL
Whole blood 13 μL
Aperture Diameter
WBC 100 μm
RBC/PLT 70 μm
Dilution Rations
Whole Blood Capillary Blood
WBC 1:304 1:293
RBC/PLT 1:44833 1:44402
Throughput
60 samples per hour
Display
Large color LCD display
Resolution: 800 x 600
Alerts
41 sets of alert message
Menu
Count, Mode, Review, Quality Control, Setup, Service,
Calibration, Help, Shutdown
Carryover
WBC, RBC, HGB < 0.5%, PLT < 1%
Input/Output
RS232x2, 1parallel printer(optional), 1 bar code scanner(optional), 1 keyboard interface
Printout
Thermal recorder, 50 mm wide paper, multitude of printout formats, printer optional
Operating Environment
Temperature 15°C~ 30°C
Humidity 30% ~ 85%
Power Requirement
AC 100V-240V, 50/60Hz
Dimension
390mm(W)x 460mm(H)x 390mm(D)
Weight
25 Kg
|