Diagnostic Audiometer
Screening Audiometer
Clinical Audiometer
PC Audiometer
AUDIOMETER - SIBELMED

SIBELSOUND 400

| Main features*: • Advanced digital technology based on DSP • Calculation of hearing loss and diagnosis • Internal database > 1000 tests • Suprathreshold tests • High frequency • Musical frequencies • Automatic masking • Connectivity with other management systems (EMR - Electronic Medical Record) • Customizable settings *see configurations" |
POSSIBLE CONFIGURATIONS
| CONFIGURATION ACCORDING TO MODEL | A |
AM |
AO |
AOM |
AOM+ |
SUPRA |
| Two channels | v |
v |
v |
v |
v |
v |
| Pure tone threshold audiometry by air conduction | v |
v |
v |
v |
v |
v |
| Pure tone threshold audiometry by bone conduction | o |
o |
v |
v |
v |
v |
| Pure tone threshold audiometry by free field | o |
o |
o |
o |
o |
o |
| Pure tone suprathreshold audiometry (SISIGRAM) | o |
o |
o |
o |
v |
v |
| Pure tone suprathreshold audiometry (Fowler-ABLB, Tone Decay, Weber, Lüscher, etc.) | o |
o |
o |
o |
o |
v |
| Speech audiometry | o |
o |
o |
o |
v |
v |
| High frequency audiometry | x |
x |
x |
o |
o |
o |
| Masking with narrow band noise | o |
v |
o |
v |
v |
v |
| Masking with white noise | o |
v |
o |
v |
v |
v |
| Masking with speech noise | o |
o |
o |
o |
v |
v |
| Synchronized masking | o |
v |
o |
v |
v |
v |
| Intermediate frequencies: 125, 750, 1500 Hz | o |
o |
o |
o |
v |
v |
| Musical frequencies | o |
o |
o |
o |
v |
v |
| Selection of exploration frequencies | v |
v |
v |
v |
v |
v |
| Continuous and pulse tone | v |
v |
v |
v |
v |
v |
| Pulse/alternating tone and modulating/alternating tone | o |
o |
o |
o |
v |
v |
| Reference tone (1 dB) | x |
x |
x |
o |
v |
v |
| Frequency modulation and amplitude modulation | x |
x |
x |
o |
o |
v |
| Calculation of hearing loss | v |
v |
v |
v |
v |
v |
| Diagnosis (COUNCIL, IMSS MEXICO, ELI, SAL, KLOCKHOFF, MOH, others...) | v |
v |
v |
v |
v |
v |
| Internal database for more than 1000 tests | o |
o |
o |
v |
v |
v |
| Intercom / Monitor | o |
o |
o |
o |
v |
v |
| USB and parallel connections to external printer | v |
v |
v |
v |
v |
v |
| USB computer connection | v |
v |
v |
v |
v |
v |
| RS232 computer connection | o |
o |
o |
o |
o |
o |
| Audiometry software (Demo) | v |
v |
v |
v |
v |
v |
| Audiometry software (Licence) | o |
o |
o |
o |
o |
v |
| Sound suppressors for air conduction | o |
o |
o |
o |
o |
o |
| External printer | o |
o |
o |
o |
o |
o |
| User manual and quick user guide | v |
v |
v |
v |
v |
v |
| Carrying bag | o |
o |
o |
o |
o |
o |
| Audiometer type according to IEC60645 | 4 |
4 |
4 |
3 |
2 |
2 |
| v = Standard | IEC60645 classification : |
| o = Optional | 4 = Screening |
| x = Not Available | 3 = Basic |
| 2 = Diagnostic |
DESCRIPTION
| The SIBELSOUND 400, is a revolutionary two-channel audiometer. The entire system is controlled by Digital Signal Processor (DSP) for reliable, quick and easy audiometric exploration of hearing thresholds and screening tests, such as suprathreshold pure tone tests. The SIBELSOUND 400 audiometer was developed by SIBEL, S.A.'s RDI department in collaboration with the Surgery Department (Ear, Nose and Throat and Audiology) of the University of Barcelona and well-known specialists in the field, in accordance with standard criteria of both national institutions such as the UNE and international institutions such as the IEC, ISO, etc. |
DSP DIGITAL TECHNOLOGY
| A Digital Signal Processor that uses an optimised microprocessor for applications that require very high-speed numerical operations. The processor is capable of working with multiple data in parallel, while its design and specific instructions are ideal for digital processing, features that distinguish a DSP from other types of processors. |
COMMUNICATION
|
One of the great qualities of the SIBELSOUND 400, is a communications
system with other equipment that allows it to: • Print directly to a USB printer • Transfer information from the unit's internal database to a PC • Communicate with a PC in real time • Export patient tests to other management systems (EMR) • Update the unit's internal firmware Using the appropriate software, communications can be established through two different channels: • USB (standard) • RS232C Serial (optional) |
SIGNALS
|
The SIBELSOUND 400 features: • Pure tones (continuous, pulse, alternate) • Frequency modulation • Amplitude modulation • Speech Noise • Narrow Band Noise • White Noise |
Audiometry software W50Patient's Card![]() Pure Tone Threshold Audiometry![]() Speech Audiometry![]() Sisi Jerger Audiometry Test![]() Fowler Audiometry Test![]() Tone Decay Audiometry Test![]() Luscher Test![]() Pure Tone HF Audiometry![]() |
|
SIBELMED W50 is a sophisticated software for viewing, storage,
transmission, analysis and introduction of audiometric tests
in a Microsoft Windows ® environment with the following
options: • Management of various patient databases • Download of audiometric test data from SIBELSOUND 400 and AC50 audiometers • Manual audiometric test data input performed with other audiometers • Display of different types of audiometric tests • Diagnosis: (COUNCIL, IMSS MEXICO, ELI, SAL, KLOCKHOFF, MOH, others...) • Choice of various types of diagnosis parameters • Test graphics • Comparison of tests from the same patient • Customizable reports • Test report printouts • Option for deleting tests from the unit |
TECHNICAL SPECIFICATIONS
Standard Frequency |
125 |
250 |
500 |
750 |
1000 |
1500 |
2000 |
3000 |
4000 |
6000 |
8000 |
Hz |
Musical Frequency |
131 |
262 |
523 |
- |
1047 |
- |
2093 |
- |
4186 |
- |
8372 |
Hz |
Air Conduction |
80 |
100 |
120 |
120 |
120 |
120 |
120 |
120 |
120 |
110 |
110 |
dB HL |
Bone Conduction |
- |
50 |
60 |
60 |
70 |
70 |
70 |
70 |
70 |
- |
- |
dB HL |
Free Field |
- |
70 |
80 |
80 |
80 |
80 |
80 |
80 |
80 |
80 |
- |
dB HL |
Masking Noise |
||||||||||||
Narrow Bandwidth AC |
60 |
80 |
100 |
100 |
100 |
100 |
100 |
100 |
100 |
100 |
90 |
dB HL |
Narrow Bandwidth BC |
- |
50 |
60 |
60 |
70 |
70 |
70 |
70 |
70 |
- |
- |
dB HL |
White AC |
100 |
dB SPL |
||||||||||
Speech |
100 |
dB SPL |
||||||||||
Logo-Audiometry |
||||||||||||
Air Conduction |
90 |
90 |
90 |
90 |
50 |
50 |
50 |
50 |
50 |
- |
- |
dB HL |
Minimum Levels |
||||||||||||
All Options |
-10 |
-10 |
-10 |
-10 |
-10 |
-10 |
-10 |
-10 |
-10 |
-10 |
-10 |
dB HL |
High Frequency |
-20 |
-20 |
-20 |
-20 |
-20 |
-20 |
-20 |
-20 |
-20 |
-20 |
-20 |
dB HL |
Level Increases |
||||||||||||
Standard |
5 |
5 |
5 |
5 |
5 |
5 |
5 |
5 |
5 |
5 |
5 |
dB |
Reference Tone |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
1 |
dB |
Frequency Precision |
± 1% |
|||||||||||
Level Precision |
± 2% |
dB SPL |
||||||||||
|
Channels: Two Functional specifications: See tables of models and frequencies Transducers: TDH39 - HDA200 - B71 (depending on the model) Screen: LCD alphanumeric backlighted 2 X 16 Protection: Time restricted high frequency signals to protect both patient and unit Self-check: Automatically checks the status of the audiometer Unit test: Allows the user or technician to check the condition of certain functions or components Custom settings: Programme to personalise the audiometer according to the needs of the user Calibrations: Equipped with calibrations according to ISO and ANSI Safety standards: EN60601-1, EN60601-1-1, EN60601-1-2, EN60601-1-4 Audiometry standards: IEC60645-1, IEC60645-2, IEC60645-4 Calibration standards: EN ISO 389-1, EN ISO 389-3, EN ISO 389-4, EN ISO 389-5, EN ISO 389-7, ANSI S3.6-2004 Working temperature: 10 to 40 ºC Relative humidity: <90% (no condensation) Power supply: 100 to 240V ±10% / 50/60 Hz ±3% Power: <50 VA Dimensions: 390 mm x 260 mm x 105 mm Weight: 2.4 kg without accessories Standard accessories: Depending on the model Optional accessories: Depending on the model |
![]() |
AS5

SPECIFICATION
Frequency |
250 |
500 |
1000 |
2000 |
3000 |
4000 |
5000 |
8000 |
Hz |
Max. Pure Tone AC |
90 |
105 |
105 |
105 |
105 |
105 |
90 |
80 |
dB HL |
Max. Pure Tone BC |
40 |
50 |
50 |
50 |
50 |
50 |
- |
- |
dB HL |
Min. Pure Tone AC/BC |
-10 |
-10 |
-10 |
-10 |
-10 |
-10 |
-10 |
-10 |
dB HL |
Level Increments |
5 |
5 |
5 |
5 |
5 |
5 |
5 |
5 |
dB HL |
Frequency Accuracy |
± 3% AT ALL FREQUENCIES |
||||||||
Levels Accuracy |
± 3 Db |
||||||||
Calibration |
ISO 389, UNE 74-020-78 |
||||||||
| Masking Noise : - White Noise - Level Range = 30 to 80 dB SPL - Level Increments = 5 dB Safety : UNE 20-613, IEC 601.1 Power Supply : 220 V 50Hz, 15 Watt Dimensions : 324mm x 246mm x 95mm |
Standard Accessories
| 1 Air Conduction Phones TDH-39 1 Bone Conduction Phone 1 Power Cord 1 Patient Response Switch 1 Audiometric Leaves (100) 1 Pen (Red, Blue) |
Audiometer - LiteratureAudiometer adalah peralatan elektronik untuk menguji pendengaran. Dalam penggunaannya, audiometer menghasilkan hanya satu nada (nada murni)pada beberapa frekuensi dan beberapa intensitas. Audiometer digunakan dalam pengujian audiometri dengan hasil berupa audiogram.
|